Catalog #24420 (1/12)
REV 5.1
OWNER’S MANUAL
CLX OnLine Residual
Chlorine Monitor
HF scientific
3170 Metro Parkway
Ft. Myers, FL 33916
Phone: 239-337-2116
Fax: 239-332-7643
Toll Free: 888-203-7248
Website: www.hfscientific.com
Vue de la page 0
1 2 3 4 5 6 ... 38 39

Résumé du contenu

Page 1

Catalog #24420 (1/12) REV 5.1 OWNER’S MANUAL CLX OnLine Residual Chlorine Monitor HF scientific 3170

Page 2

CLX (1/12) Page 5 REV 5.1 3.0 Theory of Operation The CLX has two solenoid valves, one for sample water (FLOW) and one for draining of the cuvette

Page 3 - DECLARATION OF CONFORMITY

CLX (1/12) Page 6 REV 5.1 IT MAY B E NEC ESSARY TO REMOVE THIS B ULKHEAD NUT TO ACCESS THE B OTTOM LEFT MOUNTING HOLE WITH THE B RAC KET IN THIS POS

Page 4 - Table of Contents

CLX (1/12) Page 7 REV 5.1 INLINE T-STRAINER1/4" OD PE TUBINGONLINE CHLORINE MONITORIF NECESSARY, CHANGETHE INLET FITTING HERENOTE DIRECTION OF

Page 5

CLX (1/12) Page 8 REV 5.1 High Voltage CoverCaptive ScrewPowerTerminalBlockAlarm #2TerminalBlockAlarm #1TerminalBlockRS-485TerminalBlock4-20mATermin

Page 6 - Specifications

CLX (1/12) Page 9 REV 5.1 Plugs are inserted into the RS-485 and 4-20mA cable bulkheads when shipped, to ensure a watertight seal. These plugs shou

Page 7 - 1.0 Overview

CLX (1/12) Page 10 REV 5.1 4.3.3 Relays The Alarm 1 and Alarm 2 relays are mechanical relays rated at 240 VAC 2A. Please note that the relays are l

Page 8

CLX (1/12) Page 11 REV 5.1 5.0 Operation The CLX Online Chlorine Analyzer allows for the measurement of the chlorine of process water on-line. The

Page 9

CLX (1/12) Page 12 REV 5.1 5.2 Security Access Feature The instrument is equipped with a security access code feature that can be activated in the

Page 10 - 3.0 Theory of Operation

CLX (1/12) Page 13 REV 5.1 6.0 Instrument Calibration The instrument was tested prior to leaving the factory. Since it operates from a pre-determi

Page 11

CLX (1/12) Page 14 REV 5.1 6.2 Zero (offset) Calibration Procedure Generally this calibration is only required if readings are expected to be below

Page 13

CLX (1/12) Page 15 REV 5.1 7.0 Instrument Configuration (CONFIG mode) The instrument has been designed to provide the ability to customize the ins

Page 14

CLX (1/12) Page 16 REV 5.1 The next prompt will be the chlorine level assigned to the 20MA. Select the chlorine level using the  and  buttons. Onc

Page 15

CLX (1/12) Page 17 REV 5.1 Press the  button to continue on and select the desired instrument address using the  or  buttons. Once the selection

Page 16 - 5.0 Operation

CLX (1/12) Page 18 REV 5.1 Alarm 1 Set Point: This prompt is used to select the set point for this alarm; this is indicated by “S/P” shown on the l

Page 17 - button

CLX (1/12) Page 19 REV 5.1 7.7 Units of Measurement The unit of measure can be set to either mg/L (milligrams per liter) or PPM (parts per million

Page 18

CLX (1/12) Page 20 REV 5.1 7.9 LCD Backlight Brightness The LCD backlight brightness may need to be adjusted. This is of particular interest if

Page 19 - Calibration Procedure

CLX (1/12) Page 21 REV 5.1 7.11 Cycle Time The cycle time can be changed using this menu. Please note that changing this menu will directly affect

Page 20

CLX (1/12) Page 22 REV 5.1 7.13 4mA Adjustment If the 4-20 mA setting is turned ON (7.1 Setting the 4-20 mA Output), the following two menus will a

Page 21

CLX (1/12) Page 23 REV 5.1 8.0 Additional Features and Options 8.1 Backlit LCD The backlit LCD allows for easier readability of the LCD displ

Page 22

CLX (1/12) Page 24 REV 5.1 8.2.3 Modbus Communication Modbus protocol communication manual is available HF #24569. This manual is available free o

Page 23

Catalog #24420 (1/12) REV 5.1 DECLARATION OF CONFORMITY Application of Council Directive Standards to Which Conformity is Declared: Product Safety

Page 24

CLX (1/12) Page 25 REV 5.1 9.0 Troubleshooting 9.1 CLX Fault Detection The CLX performs continuous diagnostic monitoring. In the CLX, there ar

Page 25

CLX (1/12) Page 26 REV 5.1 A level 1 fault is a system fault. This is NOT a problem that the operator can correct, and the unit must be returned to

Page 26

CLX (1/12) Page 27 REV 5.1 9.5 Diagnostic Chart Symptom Cause Cure Lower display shows MA 4-20 mA loop open Check wiring. See sections 4.3.4 and 7.

Page 27

CLX (1/12) Page 28 REV 5.1 10.0 Routine Maintenance 10.1 Maintenance Schedule The recommended schedule is shown below. It is important to repla

Page 28

CLX (1/12) Page 29 REV 5.1 After a PRIME the CLX will perform a water calibration (WCAL). It will take a few minutes to complete this procedure. 3

Page 29

CLX (1/12) Page 30 REV 5.1 INLETCHECKVALVESCAP ASSEMBLY "A"PUMPHAMMERAND SPRINGTHUMBSCREWREAGENTPUMPASSYINLETOUTLETFLOWyCHECK VALVE DETAIL

Page 30 - 9.0 Troubleshooting

CLX (1/12) Page 31 REV 5.1 10cc SyringeLeur-Lok Barbed AdapterIndicator Reagent Preparation Add about 400 ml of deionized water into the indicator b

Page 31

CLX (1/12) Page 32 REV 5.1 10.4 Cuvette Change As recommended in section 10.1 Maintenance Schedule the cuvette should be inspected at monthly inter

Page 32 - Symptom

CLX (1/12) Page 33 REV 5.1 11.0 Accessories and Replacement Parts List The items shown below are recommended accessories and replacement parts. Acc

Page 33 - 10.0 Routine Maintenance

CLX (1/12) Page 34 REV 5.1 12.0 Warranty HF scientific inc., as vendor, warrants to the original purchaser of this instrument that it will be free

Page 34

Table of Contents Section Page Specifications ...

Page 35 - AND SPRING

Table of Contents (continued) Section Page 7.6 Extended Settings...

Page 36 - Leur-Lok Barbed Adapter

CLX (1/12) Page 1 REV 5.1 Specifications Measurement Range 0 – 10.00 mg/L (PPM) Accuracy ±5% of reading or ±0.03 mg/L (PPM) whichever is greater for

Page 37

CLX (1/12) Page 2 REV 5.1 1.0 Overview The CLX Online Chlorine Analyzer allows for the reading of chlorine levels of process water on-line. The CLX

Page 38 - Accessory Catalog Number

CLX (1/12) Page 3 REV 5.1 1.2 The Display Figure 1 illustrates all the items that can appear on the display. The upper row of the display (1) is

Page 39 - 12.0 Warranty

CLX (1/12) Page 4 REV 5.1 2.0 Safety This manual contains basic instructions that must be followed during the commissioning, operation, care and m