OCM Rev. 3.9 (1/03) 15
8.0 Self-Diagnostics: System Troubleshooting
8.1 System Warning Message(s)
Automatic warning messages are generated by the OCM to provide you with specific
diagnostic information about the instrument. These are only messages and they do not
indicate failure of any component in the instrument. The following table lists the warning
messages and their associated meanings:
WARNING ASSOCIATED MEANING TYPICAL CAUSE
W01
Lamp Failure
Lamp has too low an output. The lamp
must be replaced followed by a
calibration.
W02 Calibration Failure
The wrong standard was used or there
is an internal failure.
W03 Analog Loop Failure
The 4-20mA is activated and the loop
is open.
W05
Water used for 0 ADJ is too
cloudy
The 0 ADJ water was determined to
be too cloudy for the instrument to
operate properly. Replace water and
calibrate again.
O-r Sample Over-Range
The sample presented to the OCM is
over 50 PPM.
WNG
Multiple Warning conditions
are met
A combination of the above warning
conditions has been met.
If any of the above conditions occurs, the sensor relay will automatically be set to the
alarm position.
8.2 System Failure Error
The instrument is indicating an error when “SFE” is displayed on the lower row of the
display. This is a general system failure. When this message is observed, contact the HF
scientific, inc. Technical Services department to determine a resolution to the problem.
If any of the above conditions occurs, the sensor relay will automatically be set to the
alarm position.